![Agilent Technologies 1260 Infinity II Series Скачать руководство пользователя страница 153](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-ii-series/1260-infinity-ii-series_user-manual_2867552153.webp)
Agilent InfinityLab LC Series Vialsamplers User Manual
153
Troubleshooting and Diagnostics
7
Step Commands
Step Commands
Each movement of the sampling sequence can be done under manual control.
This is useful during troubleshooting, where close observation of each of the
sampling steps is required to confirm a specific failure mode or verify
successful completion of a repair.
Each injector step command actually consists of a series of individual
commands that move the autosampler components to predefined positions,
enabling the specific step to be done.
Table 15
Injector step commands
Step
Action
Comments
Valve Bypass
Switches injection valve to the bypass
position.
Piston Home
Moves the piston to the home
position.
Needle Up
Lifts the needle arm to the upper
position.
Command also switches the
valve to bypass if it is not
already in that position.
Vial to Seat
Moves the selected vial to the seat
position.
Command also lifts the
needle to the upper position.
Needle into Sample
Lowers the needle into the sample.
Command also positions the
vial at the seat, and lifts the
needle to the upper position.
Draw
Metering device draws the defined
injection volume.
Command also positions the
vial at the seat, lifts the
needle, and lowers the needle
into vial. Command can be
done more than once
(maximum draw volume of
100 µL cannot be exceeded).
Use
Plunger Home
to reset
the metering device.
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...