Maintenance and Repair
9
Exchange the Gripper Arm
4
Rotate the arm approximately 2.5 cm (1 inch) to the left
and insert the straightened paper clip into the slit.
5
Rotate the gripper arm slowly from left to right and apply
a gentle pressure to the paper clip.
The clip will engage on an internal catch and the rotation
of the arm will be blocked.
6
Hold the paper clip in place, press the gripper release
button and rotate the gripper arm to the right.
7
The gripper arm will come off.
8
Replace the gripper arm by holding the paper clip in
place, pushing the gripper arm into the holder and
rotating the gripper arm to the left.
9
In the Local Controller close
Change gripper
.
OR
In the Agilent Lab Advisor software select
Change
gripper > Next
, click
Next
and wait until the gripper is in
the default position.
10
Select
Back
to finish the procedure.
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...