Agilent InfinityLab LC Series Vialsamplers User Manual
9
Maintenance and Repair
Exchange the Metering Seal and Piston
Exchange the Metering Seal and Piston
When
Poor injection-volume reproducibility
Leaking metering device
Tools required
p/n
Description
8710-0510
Open-end wrench 1/4 — 5/16 inch
8710-2392
4 mm Hex key
8710-2411
3 mm Hex key, 12 cm long
01018-23702 Insert
tool
Cleaning tissue and appropriate solvent like isopropanol or methanol
Parts required
p/n
Description
0905-1503
Metering seal, 100 µL
5067-5678
Piston ceramic
for 100 µL analytical head
0905-1717
Metering seal, 40 µL
5067-5920
Piston ceramic
for 40 µL analytical head
Preparations
Finish any pending acquisition job and in order to avoid leaks, stop the pump running.
N O T E
When the instrument setup has changed, configure the new setup in the online view of
the Chromatographic Data System. See
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...