Agilent InfinityLab LC Series Vialsamplers User Manual
227
Maintenance and Repair
9
Exchange the Wash Port Assembly
22
Select
Next
to move the needle into the needle seat.
23
Select
Back
to finish the procedure.
24
Close the doors.
25
Purge the wash port. Check if solvent will be delivered from the peristaltic pump.
Содержание 1260 Infinity II Series