Agilent InfinityLab LC Series Vialsamplers User Manual
203
Maintenance and Repair
9
Exchange the Sample Loop Assembly
1
In the Local Controller start the maintenance mode and
select
Change loop
function.
OR
In the Agilent Lab Advisor software select
Service &
Diagnostics
in the system screen
Maintenance
Positions > Change loop
, click
Start
and wait until the
needle assembly is in maintenance position.
2
Open the front door.
3
Unscrew the safety cover screw.
WA R N I N G
Sharp needle
Uncovered needles may cause injuries
➔
Do not touch the tip of the needle.
4
Remove the safety cover.
5
Check if the needle is in maintenance position.