Agilent InfinityLab LC Series Vialsamplers User Manual
123
Preparing the Module
5
Setting up the Vialsampler
Table 13
Control settings
Figure 22
Control settings
The Sampler control parameters are in the following sections:
•
Missing Vial
Mark the
Ignore missing vial
check box to specify that, if a vial is
missing, the injector ignores it and continues with a 6-second
dummy run. The message "Missing vial <x>" is logged, and the
system continues with the next injection.
•
Illumination
Toggles the illumination of the sample area, On or Off.
•
At Power On
The section is available when a cooler or thermostat is installed and
configured. Mark the
Turn on Thermostat
check box to specify that
the cooler/thermostat is switched on automatically when the
instrument is switched on.
•
Thermostat
The section is available when a cooler or thermostat is installed and
configured and the Constant temperature mode is selected.
Select
On
to switch on the cooler/thermostat. Specify the required
temperature in the adjacent field. The specified temperature must
be at least 5 °C below ambient for proper temperature control.
Select
Off
to switch off the cooler/thermostat.
•
Automatic Turn On
You can set a date and time at which the cooler/thermostat
switches on automatically.
•
Pump connected to Sampler
Use this section to specify the pump that is used with the
Vialsampler. If more than one pump is configured, display the
drop-down list and select the appropriate pump from the list.
•
Enable Analysis
This section is available when you have a thermostated autosampler
with a thermostat element connected and enabled only when the
Thermostat controlled by method
check box in the
Options
section
of the
Configuration
dialog box is cleared. It indicates the
temperature conditions under which an analysis will be made. Select
the
With Any Temperature
option to specify that the analysis can be
started irrespective of the temperature of the sample. Select the
Temperature /- 1°C
option to specify that the analysis can
be started only when the temperature of the sample is within 1 °C of
the specified value.
N O T E
The Enable Analysis section is disabled when Not controlled is
selected in the Temperature section.
N O T E
For additional help and support, highlight the desired area and press the
F1
key. A help
screen will open with additional information and documentation about the topic.
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...