![Agilent Technologies 1260 Infinity II Series Скачать руководство пользователя страница 176](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-ii-series/1260-infinity-ii-series_user-manual_2867552176.webp)
176
Agilent InfinityLab LC Series Vialsamplers User Manual
8
Error Information
Sampler Error Messages
Motor Temperature
Error ID: 34307
One of the motors of the transport assembly has drawn excessive current,
causing the motor to become too hot. The processor has switched OFF the
motor to prevent damage to the motor.
See figure
on page 26 for motor identification.
• Motor 0 temperature
: X-axis motor.
Motor 1 temperature
: Z-axis motor.
Motor 2 temperature
: Theta (gripper rotation) motor.
Motor 3 temperature
: Gripper motor (motor for gripper fingers).
The processor monitors the current drawn by each motor and the time the
motor is drawing current. The current drawn by the motors is dependent on
the load on each motor (friction, mass of components etc.). If the current
drawn is too high, or the time the motor draws current is too long, the error
message is generated.
Probable cause
Suggested actions
1
Mechanical obstruction.
Ensure unobstructed movement of the
transport assembly.
2
High friction in the transport assembly.
Please contact your Agilent service
representative.
3
Motor belt tension too high.
Switch OFF the autosampler at the power
switch. Wait at least 10 min before switching
on again.
4
Defective motor.
Please contact your Agilent service
representative.
5
Defective transport assembly flex board.
Please contact your Agilent service
representative.
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...