Agilent InfinityLab LC Series Vialsamplers User Manual
7
Troubleshooting and Diagnostics
Maintenance Functions
Change Gripper
The change gripper function moves the gripper to the front of the autosampler
enabling easy access to the gripper release mechanism.
User Interface
The commands for the control software are:
Start
Moves the transport assembly and gripper arm to the position required to
change the gripper arm.
Next
Repositions the transport assembly and gripper arm to the home position.
Back
Completes the procedure.
Using the
Change Gripper
Function
1
Select
Start
to move the gripper arm to the maintenance position.
2
3
Select
Next
to move the gripper arm to the home position.
4
Select
Back
to finish the gripper procedure.
N O T E
Read carefully the description in the Lab Advisor user interface.
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...