Agilent InfinityLab LC Series Vialsamplers User Manual
195
Maintenance and Repair
9
Exchange the Needle Assembly
12
Reconnect the sample-loop to the needle.
N O T E
Do not overtighten the fitting!
N O T E
Incorrect positioning and installation of the
needle/loop connection can result in damaging and
breaking the sample loop.
13
In Lab Advisor use
Next
to lift the needle slowly into the
up position.
14
Turn or flap the wash port to the right.
15
Install the safety cover (1.) and fix the screw (2.).