Agilent InfinityLab LC Series Vialsamplers User Manual
125
Preparing the Module
5
Setting up the Vialsampler
Table 14
Method parameter settings
Injection
Specify the injection volume (in µL) in the
Injection
Volume
field. The injection volume limits depend upon
the configuration settings of the autosampler. Note that
this setting is ignored when Pretreatment/Injector
Program is defined.
Needle Wash
Mark the
Enable Needle Wash
check box to specify an
injection with needle wash.
The following wash parameters can be specified:
•
Mode
: Select the wash location by clicking the down
arrow and selecting the needle wash mode from the
list. You can select from Flush Port (default location)
or Wash Vial.
•
Time
: This field is enabled only if you have selected
Flush Port. Specify a flush time in seconds between 1
and 100 in steps of 1.
•
Location
: enabled only if you have selected a Wash
Vial. Allows you to specify a wash location in the
format Vial X, where X is the vial number. Specify well
locations in the format P<Plate #>-<Row
Character><Column Number>, for example, P1-A1.
The values of <Row Character> and <Column
Number> depend on the wellplate type.
•
Repeat
: available only if you have selected Wash Vial.
You can specify up to 10 repeat washes.
Stoptime
: enables you to set the time that the analysis
stops. Limits: 0.01 through 99999 min or
As Pump/No
Limit
.
Posttime
: the Vialsampler remains in a post-run state
during the Posttime to delay the start of the next analysis.
A Posttime period can be used to allow the column to
equilibrate after changes in solvent composition (for
example after gradient elution). Limits: 0.01 through
99999 min or
Off
.
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...