Agilent InfinityLab LC Series Vialsamplers User Manual
149
Troubleshooting and Diagnostics
7
Maintenance Functions
To prepare autosampler for transportation
The park arm function moves the gripper and transport slider to the home
position. Reinstall the transport foam to secure the instrument against a
mechanical stop. The autosampler can be switched OFF after the preparation.
1
Select
Park Arm
.
2
When the arm is in the park position, install Transport-Protection
(G7129-40050). Then the autosampler is ready for shipment, and can be
switched OFF.
When
Before transporting or shipping the module.
C A U T I O N
Unsecured Transportation of the module
Unsecured transportation of the module may result in mechanical damage to the
gripper and transport slider.
➔
Always secure the arm in the park position plus transport foam.
N O T E
Before parking the gripper arm, ensure there is no vial in the gripper. Use the Release
Gripper function to remove the vial.
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...