Agilent InfinityLab LC Series Vialsamplers User Manual
1
Introduction
Leak and Waste Handling
Leak Sensor
Waste Guidance
C A U T I O N
Solvent incompatibility
The solvent DMF (dimethyl formamide) leads to corrosion of the leak sensor. The
material of the leak sensor, PVDF (polyvinylidene fluoride), is incompatible with DMF.
➔
Do not use DMF.
N O T E
The leak sensor in the sampler is hidden under the ICC Column Heater or Column Shelf
respectively.
N O T E
The waste drainage must go straight into the waste containers. The waste flow must not
be restricted at bends or joints.