Agilent InfinityLab LC Series Vialsamplers User Manual
213
Maintenance and Repair
9
Exchange the Metering Seal and Piston
1
In the Local Controller start the maintenance mode and
select
Change metering device
function.
OR
In the
Tools
section of the Agilent Lab Advisor software
select
Service & Diagnostics > Maintenance Positions
> Change Metering Device
, click
Start
and wait until
the metering device is in maintenance position.
2
Open the front door.
3
Disconnect the two capillaries from the metering device.
4
Remove the two fixing bolts.
5
Pull the metering head assembly away from the sampler.
6
Remove the two fixing bolts from the base of the
metering head assembly.
7
Disassemble the metering head assembly.
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...