Agilent InfinityLab LC Series Vialsamplers User Manual
159
Troubleshooting and Diagnostics
7
Troubleshooting Guide for the Sample Transport Assembly
Intermittent lock-ups with or without vial in the gripper fingers
With error messages
• motor overtemp
(0 or 1 or 2 or 3)
• movement failed
(0 or 1 or 2 or 3)
• missing vial
1
Check the vials and the caps.
For reliable operation, vials used with the autosampler must not have
tapered shoulders or caps that are wider than the body of the vial. For more
details, see
“List of Compatible Vials and Caps”
2
Reset the sampler and check the tension of the belts.
3
Exchange the gripper arm assembly (Gripper assembly (G1313-60010)), see
4
Start automatic referencing in Lab Advisor. For this procedure it is
mandatory to use the calibration tool.
On completion of the procedure, the corrected values are stored in the
instrument firmware.
5
Exchange Transport assembly ( Sample Transport Assembly
(G7129-60600)).
6
Exchange the Autosampler Main Board ( Autosampler Specific Board
(G7129-65800)) or the fusion board ( Fusion Board (G4200-65880)).
7
For both steps (step 5 and 6), you have to do an automatic referencing again
to store the correct values in the instrument firmware again.
WA R N I N G
Personal damage, damage to the module
➔
Some of these procedures require a trained service engineer. Persons who are not
qualified MUST NOT perform these procedures.
N O T E
When a
Motor Temperature
message has occurred, the sampler must be turned OFF for
about 10 min to allow the motor to cool down.
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...