Agilent InfinityLab LC Series Vialsamplers User Manual
115
Preparing the Module
5
Installing Capillaries
Installing Capillaries
For correct installation of capillary connections of the sampler it's important
to choose the correct fittings, see
“Capillary Color Coding Guide”
The capillaries mentioned above are examples only.
Parts required
p/n
Description
5067-4650
Capillary ST 0.12 mm x 150 mm SL/SX
5067-4651
Capillary ST 0.12 mm x 280 mm SL/SX
5067-4720
Capillary ST 0.17 mm x 150 mm SL/SX
5067-4722
Capillary ST 0.17 mm x 280 mm SL/SX
5065-4454
Fitting screw long
10/pk
Quantity depends on configuration of the module (number of connections to
the multisampler).
1
Select a nut that is long enough for the fitting you'll be
using.
2
Slide the nut over the end of the tubing or capillary.
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...