66
Agilaire 8872 Manual
Chapter 2 Configuration
Configuring Instrument Controlled Calibrations from Configuration Editors > Logger Channels
Configuring Instrument Controlled Calibrations
If you configure
Instrument Controlled
Calibrations, enabled calibrations will be initiated by
the data logger when it detects a specified digital input pattern. Each phase will continue until
the input line pattern changes. The Instrument Controlled Cal screen has the following fields:
Calibration Type
will be already filled in (InstrumentControlled_I).
Calibration Name
is required to identify the cal program.
Check
Enabled
if the calibration is to run.
Number of Calibration Records
determines how many cals the data logger will store
before overwriting.
Recovery Time
specifies the time required to purge cal gas after phases
Affected Channels
determines which channels will be taken off-line during cal.
Select from a list of previously configured parameters.