41
6. Once changed, click ‘Apply mode’.
7. In AVTrend, go to Logger Channels and select your wind speed channel.
8. On the Modbus tab, link the channel to the correct input module, and the correct input on
that module.
**NOTE: DI-0 in Apax equates to Status Inputs 1+ and 1- on the rear of the logger and to DI_1
in AVTrend; DI-1 would equate to 2+, 2- and DI_2; etc.**
9. On the Misc tab, set the High
Out and Low Out units.
To determine the values to enter use the equation: Output Value) / (Translation Value).
EX. A Met One 010C may be spec’d at a range of 0.27 to 50 meters per second, with 1250 Hz
equivalent to 50m/s.
High Out would equal (50)/(1250) Low Out would equal (0.27)/(1250)
Agilaire 8872 Manual
Chapter 2 Configuration