Calibration Alarms
Calibration alarms are set as part of the calibration configuration, using the
Alarms
tab and the
Add > Alarm
button. The user would designate:
Agilaire 8872 Manual
Chapter 2 Configuration
Calibration Alarm configuration
Alarm Name
The name to distinguish different cal alarm programs.
Enabled
Allows the alarm to be temporarily disabled.
Alarm On Cal Drift
If set, the alarm will be triggered if any of the Monitored Parameters are found to have
a difference between measured value/result and Expected Value to be greater than the
drift limit for that channel / phase.
Alarm On Cal Aborted
If set, the alarm will be triggered if the calibration sequence is aborted before it complete.
Alarm On Autoscale Failure
If set, the alarm will be triggered if “EV for Auto Correct” is set, but the data logger
cannot mathematically calculate a rescaling adjustment.
Output to Alarm Port
Not used in the Model 8872.
End Alarm on No Flag
Not used in the Model 8872.
Ack (Acknowledge) Timeout Interval
Not used in the Model 8872.
Alarm Output Lines
Used to define which relays / Modbus outputs are to be turned on when the alarm
condition is found.
Alarm Input Lines
Not used in the Model 8872.
Monitored Parameters
Used to set the list of calibration parameters that are observed for calibration purposes
(may be a subset of the total list of calibrated parameters).