Within the
Phase
settings, properties may include:
Phase Settings
Phase Name
The label used to identify this calibration stage/phase (e.g., Zero, Span, Prec)
Phase Number
Defines the order of the phases in which they run (not used for Instrument Controlled
calibrations, as their phases can occur in any order).
Duration Time (Automatic, User Initiated, Instrument Initiated)
The period of time that the control outputs for that phase will be set.
Response Time
The window of time at the end of that phase in which readings are averaged to form the
resultant value for that phase.
Status Pattern
Defines the status output or input
pattern associated with that phase.
For input patterns, the user should
define which inputs are to be observed,
and if they should be closed or opened
for that pattern.
The pattern for
each phase and calibration
should be unique.
Line Status Pattern for Phases
Agilaire 8872 Manual