68
the right side of screen or press the
key to record and continue.
Before final completion of calibration, you can return to the previous operation by using the
icon or
key anytime, until exiting the entire calibration function. This calibration will not take effect.
If the measured value deviates greatly from the calibration point, it will give a prompt.
After calibration is completed, it will take effect immediately.
3. Restore the factory calibration data.
Click
on the right side of screen, to restore the factory data.
4.4.2 Auto tune
Auto tune can optimize controller parameters which have reduced after long-term use, so as to improve the pressure
control stability and efficiency, as shown in Figure 27 to plug the output port directly before Auto tune.
Caution: Please seal the pressure outlet. Auto tune is necessary only when the pressure controlling of the unit is not
performing to specification. Incorrect or unnecessary auto tune may influence the control performance. Please carefully
use this function.