8
2. Electrical specification (Environment temperature 20
℃±
5
℃
, 1 year accuracy)
Table 4 Electrical specification
Model
Specification
Range
Resolution
Accuracy
Note
mA Measure
-
25 to 25 mA
0.1 μA
± (0.008%RD + 1.0 μA)
Impedance <10
Ω
-
50 to 50 mA
0.1 μA
± (0.008%RD + 2.0 μA)
V Measure
-
300 to 300 mV
1 μV
± (0.008%RD + 6 μV)
Impedance <1 G
Ω
V Measure
(Auto-ranging)
-
5 to 5 V
20 μV
± (0.008%RD + 100 μV)
Impedance >1 M
Ω
-
12 to 12 V
100 μV
± (0.008%RD + 320 μV)
-
30 to 30 V
100 μV
± (0.008%RD + 600 μV)
Loop Power Source
24 V
N/A
±1 V
50 mA (Max Loading)
mA Source
0 to 2.5 mA or
2.5 to 25 mA
0-2.5 mA:
0.05 μA
0 - 25 mA:
0.5 μA
0-25 mA:0.5μA
0-2.5 mA: 0.008%RD+0.2 μA
0-25 mA: 0.008%RD+1.0 μA
20 mA @ 1 K
Ω
Power Source
16 to 30 V
1 V
±1 V
70 mA (Max Loading)
V Source
0 to 16 V
0.25 mV
0.008%RD+500uV
Pressure Switch
Mechanical Switch,
Live Mechanical
Switch, NPN Switch,
PNP Switch
N/A
N/A
Response time<10
ms.
If the switch is live,
voltage range will be
(3-30) V
Temperature
Compensation
18 °C to 28°C
Temperature
Coefficient
Outside of 18 °C to 28 °C: <± (0.0005%RD + 0.00005%FS)/°C
Содержание 762
Страница 1: ...762 Automated Hydraulic Pressure Calibrator...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 18: ......
Страница 29: ...11 2 Installation 2 1 Features 2 1 1 Basic structure Figure 1 Basic structure...
Страница 31: ...13 17 USB slave port 18 Power supply 19 USB master port 20 Head correction reference line 21 Battery...
Страница 50: ...32 Figure 9 Current measurement...