105
Enable or disable loop power as required for the input.
For Enable or disable loop power as required for the output.
5. Task end
For the contents, refer to Para. 5.2.1 Task end for calibration of the dial pressure gauge.
5.2.9 Valve opening gauge
1. Connection
Connect the calibrator and the valve opening gauge as shown in Figure 32.
2. New task
Select the “valve opening gauge” icon
in the test type list and input the information of the calibrated valve
opening gauge.
Figure 32 Calibration of the valve opening gauge