Remove the filter, replace or clean the filter, and then re-press it into the corresponding groove , as shown in the
figure
③
;
4.
Check the condition of the sealing ring and replace it if it is damaged ;
5.
Insert the single connector assembly into the fixed connector, pay attention to the direction of the single connector
assembly , as shown in the figure
⑤
;
6.
Rotate the hexagonal nut clockwise until it is tightened , as shown in the figure
⑥
;
7.
Maintenance interval, 3 to 6 months/time or according to usage.
Note: Make sure to do this with the calibrator fully depressurized and powered off.
8.6 Special notice for water version
1.
Replace the water regularly (10 days is recommended), refer to Para. 8.4 for cleaning operation.
2.
If the unit is not used for a long period, it is necessary to empty the water completely including the water in the
reservoir.
3.
Minimum working temperature: 5
℃