5.1.2 Pressure transmitter
Select the “pressure transmitter” icon
in the test type list to start the quick test for pressure transmitter.
1. Connection
Refer to Para. 3.6.2 for connection.
2. Start the test
Click the zeroing button
to reset the pressure control module.
Click the start button
to select the automatic execution to enter the execution parameter setting interface.
Fill in the set point list, the number of cycles, the stroke mode, the number of readings, the reading interval, and the dwell
time, and then click the OK button
to start the test.
Selecting manual perform and then start the test. After the set point is stabilized, it is required to click the next
point button
to collect data and then input the new set point.
To stop this test, click the back button
.
3. Save the result
Click the save button
to save the test result.
The following items need to be entered: Name and serial number, click OK button
to save the test results.
To exit without saving press the back button
.