55
Figure 24 Calibration of a 4-wire pressure transmitter
3. Manually set or use the step function (see Para. 3.2.1 "Auto step and manual step of pressure output") to output the
pressure of each calibration point.
4. When the pressure of each calibration point becomes stable and is displayed in green, record the output value of the
transmitter at each point, or use the snapshot feature.
3.6.3 HART transmitter calibration
1. Prepare
Select the pressure output in menu.
Open the pressure output area of main operation interface: High-pressure or low-pressure range (depends on