110
Connect the pressure module;
First apply pressure to the leak detection point;
according to
key or the start icon on the right side of the screen to start execution, start the countdown
waiting time, and record the real-time pressure at the end of the waiting time as the starting pressure;
Start the test, count down the test time, and start to calculate the real-time leakage, real-time leakage = initial
pressure – real-time pressure;
Record the pressure at the end of the test time as the end pressure;
The whole leak detection process is over, and the final leak volume = initial pressure – ending pressure;
The whole process will be shown in stages in the leak detection curve diagram at the bottom of the screen;
4.
Check
After entering the pressure leak test interface, the previously saved pressure leak test items are displayed, click
to enter the view interface of the test
Click
and
Can switch between table and graph form
5.
delete
After entering the pressure leak test interface, click
After that, all test result items in the current interface
enter the state with selection,
You can click to select the target item, or click the icon
Check all options
Select the test result item to be deleted and click
After confirming, you can delete the selected item; click
Confirm to delete all items
Содержание 762
Страница 1: ...762 Automated Hydraulic Pressure Calibrator...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 18: ......
Страница 29: ...11 2 Installation 2 1 Features 2 1 1 Basic structure Figure 1 Basic structure...
Страница 31: ...13 17 USB slave port 18 Power supply 19 USB master port 20 Head correction reference line 21 Battery...
Страница 50: ...32 Figure 9 Current measurement...
Страница 139: ...121...