66
During pressure control, auto step or task execution, this function does not take effect.
2. Auto
sleep
In case there is no key and serial port command operations within a set time, it will automatically sleep.
There are 4 types of auto-sleep setting: never, 1 min, 5 min and 30 min.
If connected to a power adapter, the calibrator does not support auto sleep.
During pressure control, auto step or task execution, this function does not take effect.
When the backlight off time is set to "never", the auto-sleep setting is invalid.
3. Auto
power off
In case there is no key and serial port command operations within a set time, it will automatically shut down.
There are 5 types of auto power off settings: never, 5 min, 15 min, 30 min, 1 hour and 2 hours.
If connected to a power adapter, the calibrator does not support auto power off.
During pressure control, auto step or task execution, this function does not take effect.
When the backlight off time or auto sleep time is set to "never", the setting of auto power off is invalid.
4.4 System calibration
The electrical measurement, electrical output and internal pressure modules of this calibrator all need periodic
calibration.
This calibrator can also provide calibration for external pressure.
Before calibration, please carefully read the user manual, and perform operation after you have understood it.
Improper calibration will affect the accuracy of calibrator, and in severe cases, may affect normal working of the
Содержание 762
Страница 1: ...762 Automated Hydraulic Pressure Calibrator...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 18: ......
Страница 29: ...11 2 Installation 2 1 Features 2 1 1 Basic structure Figure 1 Basic structure...
Страница 31: ...13 17 USB slave port 18 Power supply 19 USB master port 20 Head correction reference line 21 Battery...
Страница 50: ...32 Figure 9 Current measurement...
Страница 139: ...121...