51
When the external pressure modules A and B are both connected, the calibrator can show up to five channels in the
multi-screen.
If the measured value exceeds the present range of pressure module, it will turn red and give an alarm.
If the measured value exceeds the measurable range of pressure module, it will display red "--------" and give an
alarm.
3. Related operations of external pressure module
The pressure unit can be switched.
Press
to enter the function menu. Select " Stability", to set the stability time and stability.
Zeroing is allowed.
Select "Module Information", to view related information of the external pressure module.
Select "Resolution", and set the number of digits displayed to 4, 5 or 6.
Select "Filter". Then, you can select first-order filter or average filter as the filter type, and set related
parameters.
3.6 Typical applications on main page
3.6.1 Pressure gauge calibration
1. Prepare
Go to the pressure output area of main operation interface:
Select High-pressure or low-pressure range (depend on the range of calibrated instrument).
2. Connection