49
6. Manual step
Press the
/
key on the keyboard or
on the screen, to manually step through values.
You can click the number inside
to set the step value of manual step.
7. Ramp
Press the
key to enter the Ramp setup interface (only 0~25mA and 0~16V is applicable). For the Ramp output
parameters, see Table 15.
Table 15 Ramp parameters
3.5 External pressure module
1. Connection: Connect the external pressure module as shown in Figure 19.
Subject
Valid value
Description
Range
Do not exceed the electrical output range
Set the ramp output range.
Rise time
1~60 sec
Ramp rise time
Fall time
1~60 sec
Ramp fall time
0%
wait time
1~60 sec
Wait time of the slope at the lower range value.
100%
wait time 1~60 sec
Wait time of the slope at the upper range value.
Repeat
0-100
Set the cycle number of slope. 0 indicates infinite repetition.