95
3. Task setting
For the contents, refer to Para. 5.2.1 Task setting for calibration of the dial pressure gauge.
4. Task start
For some contents, refer to Para. 5.2.1 Task start for calibration of the dial pressure gauge. Note that it does not
require a manual input of the indicated value in this test.
If connecting an external pressure module, it needs to be selected
If loop power is required, please set it before starting
5. For the contents of task end, refer to Para. 5.2.1 task end for calibration of the dial pressure gauge.
5.2.5 Pressure switch
1. Connection
Connect the calibrator and the pressure switch as shown in Figure 25;
2. New task
Select the “pressure switch” icon
in the test type list and input the information of the pressure switch.
Table 30 Task mode pressure switch information
Subject
Valid value
Description
Name
Letter, numeral, symbol
Name of the pressure switch
Serial number
Letter, numeral, symbol
Serial number of the pressure switch
Model
Letter, numeral, symbol
Model of the pressure switch