28
◆
Press
to go to Step mode.
◆
The set value must meet the pressure control range of the calibrator. For the pressure control range, please refer to
Para. 1. 6 Technical Specifications;
◆
When target pressure exceeds current pressure range of the internal module but within the range of the second
internal module, it will switch to the applicable module.
4. Start /stop controlling pressure
◆
The
Calibrator will start controlling after the target pressure value is achieved.
◆
On the status of vent or measure, click
or “control” button on the screen to start controlling to the target
pressure.
◆
Press
,
or click “vent” and “measure” button on the screen to stop control.
5. Pressure stabilize
◆
When stable within the control settings(see Para. 4.1), the displayed pressure value will change to green.
6. Auto step
◆
Press
to enter into auto step setting interface, see parameter in Table 10:
Table 10 Auto step settings
Subject
Valid value
Description
Stroke
Round trip or one way
Setting travel mode of auto step
Loop time
0-100
Setting loop time of auto step