107
3. Task setting
For the contents, refer to Para. 5.2.1 Task setting for calibration of the dial pressure gauge.
4. Task start
For some contents, refer to Para. 5.2.1 Task start for calibration of the dial pressure gauge.
Set loop power settings as necessary.
5. Task end
For the contents, refer to Para. 5.2.1 Task end for calibration of the dial pressure gauge.
6. Application
6.1 Unit conversion
1. Click the menu in the upper right on the main interface to select the application, and then enter the pressure unit
converter.
2. Supports conversion between multiple pressure units.
6.2 Pressure leak test
1.
Pressure leak detection and pressurization method:
Calibrator pressurization: select the internal high/low pressure range (or controllable pressure external module
A/B), and the calibrator will automatically pressurize to the leak detection pressure point;
External pressurization: Pressurized by an external pressure generator, and real-time pressure is measured by
the pressure module.