<8. Detailed Data Setting>
8-20
IM 11M12D01-01E
8.5.4 Setting Password
Unauthorized access to lower level menu displays from the Execution/Setup display can be protected
by passwords. You can set separate passwords for [Calibration, Ind. check, Blow back, and Mainte.:]
and for [Commissioning]
<Setting Procedure>
1) From the basic panel display touch the Setup key, and the [Execution/Setup] display appears.
2) Select [Commissioning] , and the Commissioning display appears.
3) Select [Others] then [Passwords]. The display shown in Figure 8.23 appears.
First set the password for [Calibration, Ind. check, Blow back, and Mainte.:]
4) Select [Calibration, Ind. check, Blow back, and Mainte.:]
5) A text entry display appears. Enter password as up to 8 alphanumeric characters.
You can enter a password for [Commissioning] by the same procedure.
6) Record the passwords and look after them carefully.
Enter
F8-23E.ai
Calibration, Ind. check
Blow back and Mainte.:
Commissioning:
Passwords
Figure 8.23 Password Setting
<Default setting>
The passwords are not set as shipped from factory. If you reset initialize the parameters, and
password settings are deleted.
CAUTION
1) If you forget the password, from the [Execution/Setup] display select [Commissioning] and enter
[JAVAJAVA] to enter Commissioning. Select [Password Setting] and confirm the password set.
2) From the Password setting display you can type spaces over an existing password to erase it.