< 4. Piping >
4-11
IM 11M12D01-01E
4.2.2 Piping for the Calibration Gases
The piping for the calibration gases should be installed between the calibration gas cylinders (or
instrument air source) and the detectors with a flowmeter, solenoid valves, needle valves, and stop
valves for performing automatic calibration.
• Place the calibration gas cylinders in a place not exposed to direct sunlight as much as possible,
e.g., house in a gas unit case. The cylinder should be equipped with a pressure reducing valve
(Yokogawa recommended part) and where necessary, a stop valve.
• Install a normally closed solenoid valve (Yokogawa recommended part) through a nipple (found on
the local market) on the calibration gas inlet of the detector as shown in Figure 4.11. This solenoid
valve is activated by a contact signal from the averaging converter. (Power wiring is separately
required.)
CAUTION
If the solenoid valve cannot be connected directly to the calibration gas inlet of the detector, install a
dedicated check valve (K9292DN or K9292DS) on the inlet and then route the piping to the solenoid
valve.
• After installing the solenoid valve on the calibration gas inlet of the detector, route the piping
through a needle valve, a flowmeter, a solenoid valve for switching zero and span gases, a stop
valve, and a pressure reducing valve for the cylinder, in this order, to the cylinder.
• Install the piping for the solenoid valve for switching zero and span gases so that the zero gas is
introduced to the detector when powered. This solenoid valve is activated by a contact signal from
the averaging converter. (Power wiring is separately required.)
• Install the needle valve and flowmeter in close proximity as much as possible.
• Use stainless steel pipes with 6 OD x 4 ID mm (or nominal 1/4 inch) or larger inside diameter for
the piping for the calibration gases.
F4-11E.ai
Solenoid valve (normal close)
Connect directly to piping inlet
Calibration gas piping
Ø6 × Ø4mm stainless steel pipe
Reference gas piping
Ø6 × Ø4mm
(or nominal 1/4-inch) stainless steel pipe
Nipple
Figure 4.11 Piping for the Calibration Gas Inlet
4.2.3 Piping for the Reference Gas
The piping for the reference gas should be installed between the air source (instrument air) and the
detector through flowmeters, needle valves and an air set for flow control of the reference gas.
• Install the piping from the reference gas inlet of the detector to the air source with a flow meter, a
needle valve where necessary, and an air set in this order.
• Install the flowmeter, needle valve and air set in close proximity.
• Use stainless steel pipes with 6 OD x 4 ID mm (or nominal 1/4 inch) or larger inside diameter for
the piping for the reference gas.