< 7. Startup >
7-9
IM 11M12D01-01E
8) Open the span gas flow setting valve, and adjust flow to 600 ± 60 ml/min.
9) Select [Valve opened (Start Calibration)] and touch the Enter key. An oxygen concentration
graph like Figure 7.13 is displayed, and [CAL.TIME] status mark flashes.
Wait for the graph measured value trace to stabilize. Calibration has not yet been performed,
so it is not a problem if the displayed value is different from the correct value of span gas
concentration.
F7-13E.ai
Enter
21.00%
Ch1
1min./div.
0.98%
21.0 %O2
CAL. TIME
Manual cal.
Channel No.
under calibration
Setting value of
span gas concentration
Setting value of
zero gas concentration
Actual measurement value
Figure 7.13 Display during calibration
10) When the graph trace stabilizes, touch the Enter key. The display changes to like Figure 7.14.
The measured value is adjusted to read the same as the span gas concentration set value, and
a check mark appears in the check box beside Span calibration.
Close the span gas flow setting valve and tighten the lock nut used to prevent it from loosening
during measurement.
Even if span calibration is executed, the span calibration will not be completed normally if a
check mark is not placed in the [Span calibration] box on the Zero calibration Start Display
shown in Figure 7.14. Even if zero calibration is executed subsequently, the results will not be
applied.
11) Next perform zero calibration. Select [Zero calibration] and touch the Enter key.
12) Select [Next] and touch the Enter key.
Enter
F7-14E.ai
Ch1
Manual cal.
Span calibration
Zero calibration
Zero calibration
End
Close the span gas valve.
F7-15E.ai
Enter
21.00%
Ch1
1min./div.
0.98%
0.98 %O2
Manual cal.
CAL. TIME
Figure 7.14 Zero calibration start display Figure 7.15 Display during calibration
13) Open the zero gas flow setting valve, and adjust flow to 600 ± 60 ml/min.
14) Select [Valve opened (Start Calibration)] and touch the Enter key. An oxygen concentration
graph like Figure 7.15 is displayed, and [CAL.TIME] status mark flashes.
Wait for the graph measured value trace to stabilize. Calibration has not yet been performed,
so it is not a problem if the displayed value is different from the correct value of zero gas
concentration.