![YOKOGAWA av550g User Manual Download Page 41](http://html.mh-extra.com/html/yokogawa/av550g/av550g_user-manual_911433041.webp)
< 2. Specifications >
2-24
IM 11M12D01-01E
2.6.6 Air Set
This set is used to lower the pressure when instrument air is used as the reference and span
gases.
Standard Specifications
• G7003XF, K9473XK
Primary Pressure:
Max. 1 MPa G
Secondary Pressure: 0.02 to 0.2 MPa G
Connection:
Rc1/4 or 1/4 FNPT (with joint adapter)
Weight:
Approx.1 kg
Part No.
Description
G7003XF
Joint: Rc 1/4, Material: Zinc alloy
K9473XK
Joint: 1/4 FNPT (with joint adapter), Material: Zinc alloy, Adapter: SUS 316
• G7004XF, K9473XG
Primary Pressure:
Max. 1 MPa G
Secondary Pressure: 0.02 to 0.5 MPa G
Connection:
Rc1/4 or 1/4 FNPT with joint adapter
Weight:
Approx. 1 kg
Part No.
Description
G7004XF
Joint: Rc 1/4, Material: Zinc alloy
K9473XG
Joint: 1/4 FNPT (with joint adapter), Material: Zinc alloy, Adapter: SUS 316
l
External Dimensions
Ø74
Primary
Secondary
88
Max. 210
max. 55
Approx. 122
Secondary
pressure
gauge
Panel (Horizontal mounting)
Panel (Vertical mounting)
G7003XF, G7004XF: Rc 1/4
K9473XK, K9473XG: 1/4 FNPT
(with joint adapter)
F2-13E.ai
A
22 ø15
2-ø2.2
+0.5
-0
2-ø6.5
40
2-ø6 screw depth 8
40
Horizontal
mounting
Vertical
mounting
Panel cut dimensions
View A
Unit : mm
Summary of Contents for av550g