< 11. Inspection and Maintenance >
11-15
IM 11M12D01-01E
11.6 Adding the Expansion Power Supply Unit
WARNING
To work safely, turn off the power to the averaging converter at the external breaker and position the
power switch to “O (OFF)” before the work.
After the work, position the power switch to “I (ON)”, attach the front cover, and then apply the power
to the averaging converter.
To connect more than 4 detectors to a 4-channel base averaging converter (AV550G-A- ), the
expansion power supply unit is required to be installed as well as addition of channel card(s)
described in Section 11.5, “Adding Channel Cards.”. Follow the procedure below to add the expansion
power supply unit.
<Procedure>
1) Before the procedure, turn off the power to the averaging converter at the external breaker.
2) Remove the front cover, and position the power switch to “O (OFF)”.
3) The power cable to be connected to the expansion power supply unit is fixed inside the
averaging converter with plastic bands (See Figure 11.11). Cut a plastic band nearest to the
connector on the end of the power cable with a nipper or relevant tool. Take care not to damage
the cable sheath.
4) Insert the connector of the power cable into the expansion power supply unit. Fix the expansion
power supply unit to the frame of the averaging converter with the 4 screws supplied. Make sure
that the connector in the rear of the expansion power supply unit is seated properly and firmly
and the screws are securely tightened.
5) Connect the heater wiring of the added detector to the corresponding heater terminals, and
attach the protection cover.
NOTE
The expansion power supply unit cannot be added in the 230 V AC version.
Cable for Expansion Power Supply
F11-11E.ai
Expansion Power Supply Unit
Figure 11.11 Adding the Expansion Power Supply Unit