![YOKOGAWA av550g User Manual Download Page 32](http://html.mh-extra.com/html/yokogawa/av550g/av550g_user-manual_911433032.webp)
< 2. Specifications >
2-15
IM 11M12D01-01E
Function block :
AI block ;
3 blocks (1 block for each channels)
Transfer the data of averaging oxygen concentration to other instruments.
DI block ;
2 blocks
Transfer the status of error and alarm to other instruments.
MAI block ;
1 block (8 channels)
Transfer the data of oxygen concentration to other instruments.
MAO block ;
1 block (8 channels)
Import the data of other instruments.
Model and Codes
Averaging Converter
Model
Suffix Code
Option Code
Description
AV550G
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
Averaging Converter
Base (*1)
-A
-B
- - - - - - - - - - - - -
- - - - - - - - - - - - -
4 Channel Base
8 Channel Base
Number of Channel
Card (*2)
-A1
-A2
-A3
-A4
-A5
-A6
-A7
-A8
-B1
-B2
-B3
-B4
-B5
-B6
-B7
-B8
- - - - - - - - - - - - -
- - - - - - - - - - - - -
- - - - - - - - - - - - -
- - - - - - - - - - - - -
- - - - - - - - - - - - -
- - - - - - - - - - - - -
- - - - - - - - - - - - -
- - - - - - - - - - - - -
- - - - - - - - - - - - -
- - - - - - - - - - - - -
- - - - - - - - - - - - -
- - - - - - - - - - - - -
- - - - - - - - - - - - -
- - - - - - - - - - - - -
- - - - - - - - - - - - -
- - - - - - - - - - - - -
1 Oxygen Channel Card, Common Isolation
2 Oxygen Channel Cards, Common Isolation
3 Oxygen Channel Cards, Common Isolation
4 Oxygen Channel Cards, Common Isolation
5 Oxygen Channel Cards, Common Isolation
6 Oxygen Channel Cards, Common Isolation
7 Oxygen Channel Cards, Common Isolation
8 Oxygen Channel Cards, Common Isolation
1 Oxygen Channel Card, Individual Isolation
2 Oxygen Channel Cards, Individual Isolation
3 Oxygen Channel Cards, Individual Isolation
4 Oxygen Channel Cards, Individual Isolation
5 Oxygen Channel Cards, Individual Isolation
6 Oxygen Channel Cards, Individual Isolation
7 Oxygen Channel Cards, Individual Isolation
8 Oxygen Channel Cards, Individual Isolation
Display
-J
-E
-F
-G
- - - - - - - - - - - - -
- - - - - - - - - - - - -
- - - - - - - - - - - - -
- - - - - - - - - - - - -
Japanese
English
French
German
Power supply
-1
-2
- - - - - - - - - - - - -
- - - - - - - - - - - - -
100 / 115 V AC
230 V AC (*3)
Communication
-E
-F
- - - - - - - - - - - - -
- - - - - - - - - - - - -
HART communication
FOUNDATION Fieldbus communication (*4)
Options
/SCT
- - - - - - - -
/24
- - - - - - - - - -
/G
- - - - - - -
Stainless steel tag plate
24 Voltage output for Solenoid valve
Cable gland (Numbers in
(*5)
(*1) Select code “-B” (8 Channel Base) when future expansion exceeding 4 channels is expected.
By so doing, the expansion can be made economically.
(*2) Common isolation is recommended, when the same instrument receives the analog outputs from each channel card.
Individual isolation is recommended to prevent the trouble by mutual interference, when different instrument receives
the analog outputs from each channel card.
(*3) When suffix code “-2” (230 V AC) is selected, select code “-A” (4 Channel Base).
(*4) When suffix code “-F” (FOUNDATION Fieldbus communication) is selected, used exclusively for communication.
When using AV550G as CE marking compliance product, select HART communication.
(*5) Input 01 to 30 in
.
Standard Accessories
Name
Part No
Quantity
Remarks
Fuse
A1112EF
2
2.5 A
Hexagonal Allen Wrench
L9827AS
1
For lock screw