< 2. Specifications >
2-20
IM 11M12D01-01E
2.6 Other Equipments
2.6.1 Dust Filter for the Detector (K9471UA)
This filter is used to protect the detector sensor from a corrosive dust components or high velocity
dust in recovery boilers and cement kilns.
Sample gas flow rate is needed to be 1 m/sec or more to replace gas inside zirconia sensor.
• Standard specification
Applicable detector:
Standard-type detector for general-purpose (the sample gas flow should
be approximately perpendicular to the probe.)
Material:
Carborundum (Filter), SUS316 (JIS)
Mesh:
30 microns
Weight:
Approx. 0.2 kg
Part No.
Description
K9471UA
Filter
K9471UX
Tool
Ø51
32
10
Carborundum filter (SiC)
Detector
Unit : mm
Screw
Increasing of insertion length
F2-7E.ai
Attach the filter unit to the tip of
the detector by screwing it clockwise.
2.6.2 Dust Guard Protector (K9471UC)
Recommended to be used when sample gas is likely to flow directly into the cell due to its flow
direction in the stack or the like, flammable dust may go into the cell, or water drops are likely to fall
and remain in the cell during downtime or the like due to the installation position.
Material:
SUS316(JIS)
Weight:
Approx. 0.3 kg
4-
Ø6
122
Ø50.8
100
Increasing of insertion length
F2-8E.ai