![YOKOGAWA av550g User Manual Download Page 187](http://html.mh-extra.com/html/yokogawa/av550g/av550g_user-manual_911433187.webp)
< 12. Troubleshooting >
12-2
IM 11M12D01-01E
When a system error occurs:
1) The power to the heaters of all detectors is removed.
2) The averaging outputs and all individual outputs are put into the status specified in the [Output
hold setup] display.
3) An error icon flashes in the switch indication area.
4) The red lamps turn on in the status display lamps on all channels.
5) The individual error contacts for all channels and common error contact are activated.
12.1.3 Error Displays
By touching the error icon when an error occurs, detailed error descriptions can be displayed. Use
the Next key to move to the next page if the error descriptions are displayed in multiple pages. On
the detailed error description display, touch the Update key to update the error occurrence condition.
When an error and an alarm occur at a time, an error icon takes precedence.
Tag:
Ch1
Err2
Ch2
0.96
%
Ch3
0.96
%
Ch4
0.96
%
Ch5
0.96
%
Ch6
Err1
Ch7
0.96
%
Ch8
0.94
%
F12-1E.ai
Hold
Channels Where Errors Occur (displayed in red)
Error Icon
Tag:
Ch1
Err2
Ave-a 0.96 %O2
Ave-b 0.95 %O2
Ave-c 0.96 %O2
Hold
F12-2E.ai
Figure 12.1 Error Displays on
Multi-channel Display
Figure 12.2 Error Displays on
Single-channel Display
Tag:
Err4 : Ctrl. card
Ch1 Err1 : Cell voltage
Ch2 Err2 : Heater temp.
Ch3 Err2 : Heater temp.
Ch4 Err1 : Cell voltage
Ch5 Err1 : Cell voltage
Ch6 Err2 : Heater temp.
Ch7 Err2 : Heater temp.
F12-3E.ai
Next
Tag:
Ch8 Err3 : Ch. card
F12-4E.ai
Previous
Figure 12.3 Error Descriptions (page 1)
Figure 12.4 Error Descriptions (page 2)