< 7. Startup >
7-12
IM 11M12D01-01E
7.11.2 Checking contacts used to operate solenoid valves during automatic
calibration
1) From the Contact check display of Figure 7.18 select [Solenoid valve contacts]. From the
corresponding Solenoid valve contacts select display of Figure 7.20 select the solenoid to check.
To check the automatic calibration gas switching valve, select [Com].
Touching the Enter key opens the [Span] (Span valve open) / [Zero] (Zero valve open) selection
window. Selecting an output and touching Enter key allows you to test Open / Close the
corresponding solenoid valve.
2) To check a channel switching valve, select its channel [Ch1] through [Ch8] and touch the Enter
key to open the [OFF] (valve closed) / [ON] (valve open) selection window.
Selecting an output and touching Enter key allows you to test Open / Close the corresponding
solenoid valve.
NOTE
You cannot check a Channel Select valve if the corresponding Channel unit is not installed.
Enter
Ch1 : Zero
Com:
Ch2 : OFF
Ch3 : OFF
Ch4 : OFF
Ch5 : OFF
Ch6 : OFF
Ch7 : OFF
Ch8 : OFF
F7-20E.ai
Solenoid valve contacts
Span
Figure 7.20 Solenoid valve contact selects
7.11.3 Checking Contact Inputs
1) From the Contact check display of Figure 7.18 select [Input contacts] then touch the Enter key.
The display of Figure 7.21 appears.
2) The input contact status [Open] or [Closed] is displayed in real time.
Enter
F7-21E.ai
Contact 1 : Open
Contact 2 : Open
Input contacts
Figure 7.21 Contact input check