• If the temperature of a sample gas is high and its pressure exceeds 0.49 kPa, the temperature
of the sample gas at the detector may not be below 700°C. In this case, connect a needle valve
(found on the local market) through a nipple (found on the local market) to the sample gas outlet
(Rc1/2) of the probe adapter so that the sample gas exhaust volume can be restricted. In cases
where condensation is likely to occur in the probe adapter when the sample gas is cooled, protect
the probe adapter with an insulating material as shown in Figure 4.5.
Reducing nipple
(R1/2-R1/4 or R1/2-1/4 NPT)
Needle valve
(Rc1/4 or 1/4 NPT)
F4-4E.ai
Figure 4.4 Mounting the needle valve for restricting exhaust flow of the sample gas
Sample gas outlet
Cover flange
High temperature probe adapter
Detector
Insulating material
Probe
F4-5E.ai
Figure 4.5 Preventing to condensation