< 3. Installation >
3-9
IM 11M12D01-01E
(9) The converter should be near the detectors: the two-way resistance between detector and
converter should not exceed 10Ω.
WARNING
The averaging converter is heavy. Take extra care not to accidentally drop it. Installation work should
be done by an adequate number of people.
CAUTION
The averaging converter is heavy (100/115 V AC version: approx. 13 kg, 230 V AC version: approx. 25
kg) so needs to be installed on a wall or construction strong enough to support.
3.3.2 Installation in an Instrument Panel
When installing the averaging converter in an instrument panel, operability, maintainability, and
environmental resistance should be considered.
(1) Installation Space and Relationship to Peripheral Devices
Follow the instructions below to determine the positions of the averaging converter and peripheral
devices.
• Do not place heat-generating devices in close proximity of the averaging converter:
specially, avoid a place under the averaging converter.
• Do not place power related devices such as transformers or electromagnetic switches in close
proximity of the averaging converter.
• Segregate wiring ducts for signal lines and those for power lines and do not place in close
proximity.
Figure 3.8 shows the requirements for the mounting holes for the averaging converter.
See Section 2.4 for the dimensions of the averaging converter and maintenance clearance.
465
150
165
6-M5 screws
Unit: mm
F3-9E.ai
300 267
Figure 3.8 Mounting holes