< 9. Calibration >
9-7
IM 11M12D01-01E
Ch1 Span CAL
CAL time
CAL start
Off
On
On
Off
Zero gas ON
Span gas on
Zero gas ON
Span gas ON
CAL start in
F9-6E.ai
Ch1 Zero CAL
Ch1 Output hold
Ch2 Span CAL
Ch2 Zero CAL
Ch2 Output hold
Average output
SV-Ch1 contact
SV-Ch2 contact
SV-Ch2 contact
Ch2 Output hold
CAL time
Ch1 Output hold
CAL time
Stabilization
time
CAL time
Ch1 removed from average
Ch2 removed from average
Stabilization
time
Figure 9.6 Calibration Timing
<Setting procedure>
1) From the [Calibration setup] display, select [Timing] and the display shown in Figure 9.7 appears.
In this example, [Mode] is set to [Auto].
2) Select items and enter numerical values as required.
Enter
F9-7E.ai
Hold time: 2 min 0 0 s
Cal. time: 2 min 0 0 s
lnterval: 1 d 0 0 h
Start date: 2 2 / 0 6 / 0 4
Start time: 1 1 : 0 0
Calibration timing
Figure 9.7 Calibration time setting