< 11. Inspection and Maintenance >
11-3
IM 11M12D01-01E
Channel Card
CELL, TC, CJ
F11-2E.ai
Detector
Power supply wire for Heater
Averaging Converter
Hot Swap Switch Mechanical Relays
Power Supply Circuit
Figure 11.2
<Procedure>
1) On the basic panel display touch the Setup key. The [Execution/Setup] display will appear.
2) Use the Cursor key to move the cursor to Maintenance. Touch the Enter key.
3) On the [Maintenance] display, select [Channel card power]. Touch the Enter key. A display shown
in Figure 11.3 will appear.
4) On the [Channel card power] display, select the channel of a detector to be disconnected for
inspection or maintenance. Touch the Enter key.
5) A dropdown menu (Disable/Enable) will appear. Select [Disable] and touch the Enter key. Make
sure that all status lamp(s) of disabled channel(s) have turned off.
Enter
Ch1 : ZR22G
Ch2 : Enable
Ch3 : Enable
Ch4 : Enable
Ch5 : Enable
Ch6 : Enable
Ch7 : Enable
Ch8 : Enable
F11-3E.ai
Disable
Channel card power
Figure 11.3 Channel Card Power (Hot Swap) Display
WARNING
Before the inspection or maintenance of a detector, make sure that the power has been turned off
by measuring the voltage of the heater terminal in the terminal box of the detector for more than two
seconds with a voltmeter.
NOTE
The state (Enable/Disable) of the hot swap function is stored even after the power switch of the
averaging converter is turned off. When the power switch of the averaging converter is turned on,
power is not supplied to the disabled channel(s).