< 10. Other Functions >
10-23
IM 11M12D01-01E
<Default setting>
Blow back parameters are set as shown in Table 10.6 prior factory shipment or after data initialization.
Table 10.6 Blow back Default Setting
Item
Default
Mode
No function(disabled)
Hold time
3 minutes 00 second
Blow back time
3 minutes 00 second
Blow back interval
30 day 00 hour
Start date
01/01/01
Start time
00: 00
10.5.5 Performing Blow back
Semi-Auto Blow back
Semi-auto blow back can be started from the display or by an external contact input.
<Procedure>
Starting blow back from the display
1) On the basic panel display, touch the Setup key to display the [Execution/Setup] display. Select
[Blow back]. Next select [Start blow back] and touch the Enter key. Blow back is performed
starting from Channel 1.
2) If a display like Figure 10.27 appears, then mode is set to [No function]. Refer to Subsection
10.5.1, “Mode” and change the mode appropriately.
3) During blow back a display like Figure 10.28 is displayed, with the affected channel number
blinking.
4) To abort blow back, you can touch the Reject key to skip the blow back time and enter the hold
time. If you touch the Enter key here, then blow back for the next channel starts; if you touch the
Reject key, then hold time is skipped and you revert to measurement mode.
Enter
F10-27E.ai
Blow back
Blow back mode is
No function
or Not set contact.
Ch1
9. 1 9 %
O2
Ch2
2 1. 9 %
O2
Ch3
2 0. 3 %
O2
Ch4
2 1. 7 %
O2
Ch5
2 1. 1 %
O2
Ch6
2 0. 7 %
O2
Ch7
2 0. 3 %
O2
Ch8
1 9. 6 %
O2
F10-28E.ai
BLOW TIME
Blow back
Figure 10.27 No Function (Disabled) Figure 10.28 Performing blow back