< 4. Piping >
4-3
IM 11M12D01-01E
4.1.2 Piping for the Calibration Gas Inlet
This piping is to be installed between the zero gas cylinder and the ZA8F flow setting unit, and
between the ZA8F flow setting unit and the ZR22G detector.
The cylinder should be placed in a calibration gas unit case or the like to avoid any direct sunlight
or radiant heat so that the gas cylinder temperature does not exceed 40°C. Mount a regulator valve
(specified by Yokogawa) on the cylinder.
Mount a check valve or stop valve (specified by Yokogawa) on the nipple (found on the local market)
at the calibration gas inlet of the detector as illustrated in Figure 4.2.
(The check valve or the stop valve may have been mounted on the detector when shipped.)
Connect the flow setting unit and the detector to a stainless steel pipe 6 mm (O.D.) x 4 mm or larger
(I.D.) (or nominal size 1/4 inch).
F4-2E.ai
Stop valve or Check valve
Piping for the calibration gas, 6 mm (O.D.)
by 4 mm (I.D.) Stainless steel pipe
Piping for the reference gas, 6 mm (O.D.)
by 4 mm (I.D.)Stainless steel pipe
Figure 4.2 Piping for the calibration gas inlet
4.1.3 Piping for the Reference Gas Inlet
Reference gas piping is required between the air source (instrument air) and the ZA8F flow setting
unit, and between the ZA8F flow setting unit and the ZR22G detector.
Insert the air set next to the flow setting unit in the piping between the air source and the flow setting unit.
Use a 6 mm (O.D.) x 4 mm or larger (I.D.) (or nominal size 1/4 inch) stainless steel pipe between the
flow setting unit and the detector.
4.1.4 Piping to the High Temperature Probe Adapter
A sample gas should be at a temperature below 700
°
C before it reaches the detector. Also, a sample
gas needs to be sucked if its pressure is negative. The piping for these purposes should be connected
to the high temperature probe adapter. For the use of the high temperature probe adapter when the
high temperature detector is used, refer to Subsection 3.2.2.
• If the pressure of a sample gas is negative, connect an ejector assembly (E7046EC/ E7046EN) as
shown in Figure 4.3. Mount the pressure gauge as close as possible to the ejector. However, if the
ambient temperature is too high, mount the gauge in a location with a temperature below 40°C.
Ejector assembly for
high temperature
High temperature
probe adapter
Detector
Pressure gauge
Ejector
F4-3E.ai
Figure 4.3 Mounting the ejector assembly