XT2640 Operating Manual
13 July 2016
Page 154 of 187
COMMAND
KEYWORD
FIELD(s)
FIELD
FORMAT
FIELD
DATA
RANGE
DESCRIPTION
SCOPEVIEW?
c
CDEF
‐
Responds with
n
data points of scope view data defined by
c
and
s
, starting at time
start
and ending at
end
(both in seconds relative to the trigger detection time)
Each data point is three fields as follows –
1
st
field: NR1 indicating if this data point contains data (1) or not (0)
2
nd
field: NR3 indicating the minimum level at this data point
3
rd
field: NR3 indicating the maximum level at this data point
Note – the response can be a very large number of characters (up to 55.3K) and
fields (up to 6144).
Note – if the requested timespan is a small fraction of the captured timespan then
not all response data may have valid data.
s
KEYWORD
V or A or W
n
NR1
2 to 2048
start
NR3
‐
end
NR3
>
start
STBYERR?
c
CDEF
‐
Responds with a NR1 indicating if the XT2640 measurement error was outside of
the requirements of EN50564 at any time during the measurement (responds with
1) or not (responds with 0)
STBYHISTORY?
c
CDEF
‐
Responds with
n
data points of standby historical data defined by
def
, ending at
relative time
end
(in seconds relative to the time at which the standby
measurement was started)
The
def
field is‐
W: responds with the Watts data
V: responds with the Volts data
VCF: responds with the Volts Crest Factor data
VTHD: responds with the Volts Distortion data (in percent)
FREQ: responds with the Frequency data
A: responds with the Amps data
ACF: responds with the Amps Crest Factor data
PF: responds with the Power Factor data
ERR: responds with the XT2640 Measurement Error data (in Watts)
Each data point is three fields as follows –
1
st
field: NR1 indicating if this data point contains data (1) or not (0)
2
nd
field: NR3 indicating the maximum data recorded
3
rd
field: NR3 indicating the minimum data recorded
Note – the response can be a very large number of characters (up to 24.6K) and
fields (up to 3072).
n
NR1
2 to 1024
end
NR3
>
0.0
def
STRING
W, V, VCF,
VTHD, FREQ,
A, ACF, PF or
ERR
STBYVCF?
c
CDEF
‐
Responds with a NR1 indicating if the voltage crest factor was outside of the
requirements of EN50564 at any time during the measurement (responds with 1)
or not (responds with 0)
STBYVTHD?
c
CDEF
‐
Responds with a NR1 indicating if the voltage distortion was outside of the
requirements of EN50564 at any time during the measurement (responds with 1)
or not (responds with 0)
24.5
RDEF
MEASUREMENT
DEFINITION
FIELD
SUB
‐
FIELDS
This field is made up of one to five sub‐fields, each separated by the sub‐field separator (colon character). The sub‐fields may be
specified in any order and may optionally be omitted (a default value being used as needed).
Sub
‐
Field
Sub
‐
field
Format
Value
Description
Measurement Data
KEYWORD
FREQ
Responds with a signal frequency (in Hz)
PERIOD
Responds with a signal period (in seconds)
INTEGTIME
Responds with the integration time (in Hours)
VOLTS or V
Responds with a voltage (in V)
VPH‐PH
Responds with an inter‐phase voltage (in V). The measurement
source field must be CH1 through CH4 defining a channel which is
configured in a VPA which is configured for 3p3w(3ch) or 3p4w
wiring.
AMPS or A
Responds with a current (in A)
WATTS or W
DEFAULT if no other Measurement Data sub‐field found
Responds with a real power (in W)
LOSS
Responds with a real power loss (in W)
EFFICIENCY or EFF Responds with a real power efficiency (in %)
VAR
Responds with an imaginary power (in W)
VA
Responds with an apparent power (in VA)
PF
Responds with power factor
PHASE
Responds with apparent phase (cos
‐1
(PF), in degrees)
LOADZ or ZLOAD
Responds with load impedance (in ohms)
SERIESR
Responds with load series resistance (in ohms)
SERIESL
Responds with load series inductance (in henries)
PARALLELR
Responds with load parallel resistance (in ohms)
PARALLELC
Responds with load parallel capacitance (in farads)
SPEED
Responds with motor speed (in rpm)
SLIP
Responds with motor slip (in %)
TORQUE
Responds with motor torque (in Nm)
HP
Responds with motor power (in HP, use WATTS or W for motor
power in watts)
STBYTIME
Responds with elapsed standby power measurement time (in hours)