XT2640 Operating Manual
13 July 2016
Page 147 of 187
24.4.9.4
SCOPE
VIEW
CONFIGURATION
COMMANDS
COMMAND
KEYWORD
FIELD(s)
FIELD
FORMAT
FIELD
DATA
RANGE
DESCRIPTION
SCOPEINRUSH
e
NR1
0 or 1
Sets if performing a scope capture will temporarily override the A RANGE setting to
HI and the BANDWIDTH setting to UNFILTERED (
e
=1) or not (
e
=0).
TIMEBASE
b
NR1
0 to 20
Sets the scope data capture timebase to 5µs (
b
=0), 10µs (
b
=1), 20µs (
b
=2), 50µs
(
b
=3), 100µs (
b
=4), 200µs (
b
=5), 500µs (
b
=6), 1ms (
b
=7), 2ms (
b
=8), 5ms (
b
=9),
10ms (
b
=10), 20ms (
b
=11), 50ms (
b
=12), 100ms (
b
=13), 200ms (
b
=14), 500ms
(
b
=15), 1s (
b
=16), 2s (
b
=17), 5s (
b
=18), 10s (
b
=19) or 20s (
b
=20)
TRIGGER
i
NR1
0 to 7
Configures scope data capture triggering
Trigger Input (
i
) :
i
=0: Trigger using CH 1 V input signal
i
=1: Trigger using CH 1 A input signal
i
=2: Trigger using CH 2 V input signal
i
=3: Trigger using CH 2 A input signal
i
=4: Trigger using CH 3 V input signal
i
=5: Trigger using CH 3 A input signal
i
=6: Trigger using CH 4 V input signal
i
=7: Trigger using CH 4 A input signal
Trigger Detection Method (
t
) :
t
=0: DC rising edge
t
=1: DC falling edge
t
=2: Rectified signal (rising edge)
t
=3: High Frequency
Trigger Position (
p
) :
p
=0: Trigger is at 0% of the captured time span
p
=1: Trigger is at 25% of the captured time span
p
=2: Trigger is at 50% of the captured time span
p
=3: Trigger is at 75% of the captured time span
Trigger Level (
l
) in V or A units as applicable
t
NR1
0 to 3
p
NR1
0 to 3
l
NR3
Any
24.4.9.5
INTEGRATION
CONFIGURATION
COMMANDS
COMMAND
KEYWORD
FIELD(s)
FIELD
FORMAT
FIELD
DATA
RANGE
DESCRIPTION
INTEGAUTO
i
NR1
0 or 1
Sets that integration will be automatically timed (
i
=1) or manually
stopped and started (
i
=0)
INTEGDELAY
d
NR1
0 to 99
Sets the automatically timed integration start delay time to
d
days,
h
hours,
m
minutes plus
s
seconds.
h
NR1
0 to 99
m
NR1
0 to 99
s
NR1
0 to 99
INTEGRUN
d
NR1
0 to 99
Sets the automatically timed integration run time to
d
days,
h
hours,
m
minutes plus
s
seconds.
h
NR1
0 to 99
m
NR1
0 to 99
s
NR1
0 to 99
24.4.9.6
MOTOR
MEASUREMENT
CONFIGURATION
COMMANDS
These commands only have an effect if a motor channel is installed, but they are still valid if no motor channel is installed.
COMMAND
KEYWORD
FIELD(s)
FIELD
FORMAT
FIELD
DATA
RANGE
DESCRIPTION
DIR
n
NR1
0 or 1
Sets that the DIR input is taken as meaning a forward direction if HI (
n
=0) or
LO (
n
=1) at the time of the detected SPD input edge. This only has effect for a
SPD setting of 3.
MTRDIGITS
n
NR1
3 to 6
Sets that motor results shall be displayed with
n
digits on the POWER DATA
and CUSTOM screens.
MTREFFGROUP
n
NR1
0 to 3
Sets the efficiency group for motor power to None (
n
=0), IN (
n
=1), MIDDLE
(
n
=2) or OUT (
n
=3)
MTRPERIOD
n
NR1
0 to 11
Sets the measurement period for motor measurement according to the value of
n
as follows‐
0: 10ms measurement period
1: 30ms measurement period
2: 100ms measurement period
3: 300ms measurement period
4: 1s measurement period
5: 3s measurement period
6: 10s measurement period
7: 30s measurement period
8: 100s measurement period
9: Measurement period synchronous with VPA1
10: Measurement period synchronous with VPA2
11: Measurement period synchronous with VPA3