XT2640 Operating Manual
13 July 2016
Page 166 of 187
28
ELECTRICAL
CHANNEL
INPUT
AND
ACCURACY
SPECIFICATIONS
Specifications are valid under the following conditions‐
All specifications are valid following a 20 minute warm‐up period after turning power ON in the XT2640, when operated
from the specified source of power and within the specified environmental conditions.
All specifications in the tables of this section other than those labelled Base Scaling Error are valid for the lifetime of the
XT2640; Base Scaling Error specifications are valid for up to 2 years after calibration in normal use, or 6 months when
continuously used above 75% of the maximum specified voltage or current signal range.
All specifications are valid at the XT2640 terminals.
DC floor specifications assume that AUTOZERO is ON. If AUTOZERO is OFF add the DC Floor specification per C from the
temperature of the last performed INT DC ZERO or when AUTOZERO was last ON (whichever occurred last).
MAINS specifications are valid for signals with 45Hz to 65Hz fundamental with 20 to 100 harmonics configured and when
using AUTO‐TRACK bandwidth limiting; otherwise use the AVIONICS or AC specifications as applicable.
AVIONICS specifications are valid for signals with 300Hz to 900Hz fundamental with 20 to 62 harmonics configured or for
signals with 45Hz to 65Hz fundamental with >100 harmonics configured, and when using AUTO‐TRACK bandwidth
limiting; otherwise use the AC or MAINS specifications.
Accuracy specifications are valid for Crest Factors <100 (within peak measurable input signal range and bandwidth
limitations).
Harmonics specifications are valid for <(10%/configured number of harmonics) cycle‐to‐cycle jitter in fundamental
frequency and with each harmonic <(100%/harmonic number) of the total signal.
Note:
The No Damage input signal range is that which will not cause immediate damage. Continued use of these levels may
reduce reliability and/or future accuracy.
The Measurable input signal range is that which can be measured (the No Damage limits may limit the time for which they
may be applied). The maximum measurable level shown is a typical value; the actual measurable maximum level is within
±2% of that shown.
The Specified input signal range is that of continuously applied signal levels for which measurements are guaranteed to be
within the specified accuracies. The RMS level is also the maximum rated working signal level for safety purposes.
Accuracy specifications are guaranteed maximum errors. You should round the total maximum error upwards to the next
integer count of resolution (e.g. if the total error is 18.3mV and the data is read with 1mV resolution then use 19mV as the
maximum error).
All percentages are % of reading unless otherwise described.
When a signal has multiple significant frequency components (e.g. significant DC and AC components), add the relevant
specifications for each such component.
V
DC
, A
DC
, V
AC
, A
AC
, V
AC+DC
, A
AC+DC
, V
RDG
, A
RDG
etc. indicate the relevant voltage, current etc. reading.
H is the harmonic or spectrum point number.
N is the configured number of harmonics or spectrum points.
F or F
FUND
is the frequency of the signal in kHz.
F
BW
is the frequency of the user bandwidth setting in kHz.
28.1
INPUT
ISOLATION
SPECIFICATIONS
Valid for any V terminal to XT2640 chassis ground; any A terminal to XT2640 chassis ground; and between any V and any A terminal.
Impedance
>1GΩ || <30pF
Max. Voltage
4500V
PK
max without damage
2500V
RMS
max for <1s without damage
1000V
RMS
max continuous rated working voltage (CAT I/II)
600V
RMS
max continuous rated working voltage (CAT III)
300V
RMS
max continuous rated working voltage (CAT IV)
28.2
VOLTAGE
MEASUREMENT
SPECIFICATIONS
The specifications for voltage are independent of the current input option installed in the respective channel.
28.2.1
VOLTAGE
INPUT
CAPABILITY
AND
CHARACTERISTICS
Specification
S
Channel
Type
A
Channel
Type
L
Channel
Type
W
Channel
Type
No
Damage
Voltage
Range
<1ms
<3000V
RMS
and V
PK
<500V
RMS
and 3000V
PK
<3000V
RMS
and V
PK
<100ms
<2000V
RMS
<300V
RMS
<1500V
RMS
<5s
<1500V
RMS
<250V
RMS
<1000V
RMS
Continuous
<1000V
RMS
<160V
RMS
<650V
RMS
XT2640 Unpowered
As above
Measurable
Voltage
Range
<1803V
RMS
and V
PK
<182.3V
RMS
and V
PK
<1803V
RMS
and V
PK
Specified
Voltage
Range
<1000V
RMS
and <1750V
PK
<160V
RMS
and <175V
PK
<650V
RMS
and <1750V
PK
Impedance
Burden
1.201MΩ ± 0.25%
121kΩ ± 0.25%
399.5kΩ ± 0.25%
3dB
Bandwidth
(typical)
900kHz
3MHz